Pump
¸ðµ¨ |
GilAir
5 |
Total
Flow Range |
1 -5,000
cc/min* |
High
Flow Range
(flow
control ¡¾5% of set
point,
except model BDX-II) |
750
- 5,000 cc/min
0.75-5.0
LPM |
Low
Flow Range,
Constant
Flow
(flow
control ¡¾5% of set point) |
1-500cc/min,
to 25"
(640mm)H20
* |
Low
Flow Range,
Constant
Pressure(Multi Flow) |
1-750cc/min,
at 18"¡¾3" H20
*
(460¡¾80mm) |
Max.
Pressure Capability: |
Fault¢Ó |
8hr.
Run¢Ó¢Ó |
@ 0.75
LPM,inches H20 |
35 |
20 |
@ 1
LPM,inches H20 |
37 |
29 |
@ 2
LPM,inches H20 |
37 |
26 |
@ 2.5
LPM,inches H20 |
- |
- |
@ 3
LPM,inches H20 |
32 |
23 |
@ 4
LPM,inches H20 |
20 |
18 |
@ 5
LPM,inches H20 |
10 |
8 |
Data
Shown In Display |
Elapsed
Time(clock & timer models),
Low
Battery,
Flow
Fault |
Overall
Dimensions |
3.6"W
x 3.9"H x 2.0"D
9.1cmW
x 9.9cmH x 5.1cmD |
Total
Pump Weight |
22.5
oz. (0.64kg) |
Rechargeable
Nicad Battery |
6.0v,
1.8amp.hr. |
Operating
Temperature Limitations |
32¢µ
to 104¢µ
0¡É to
40¡É |
Certifications |
CE,Ex,UL:
Class 1, Groups A,B,C,D;Class II,
E,F,G;and
Class III; Div. 1; Temp. Code T3C **** |
Charging
Time |
14 ~
16 hours |
Battery
Life ** |
300-500charge-recharge
cycle,
OR 2.5
yrs with<20hrs.weekly use,
1.5-2.5
yrs(20-39 hrs. weekly use),
1.0-1.5
yrs(40-60 hrs.weekly use) |
Operating
Duration (full charge) |
8 to
10 hrs. at 3.0LPM(see back pressure, above)
(except
BDX II, rated at 10 hrs. @ 2.0 LPM) |
* 1-750cc/min. requires
installation of low flow module.
** Battery Life estimates
based on proper battery maintenance.
*** Available in models
HFS-513-AUC or HFS-513-AUP only.
**** Certifications can
vary, based on model.
¢Ó Maximum backpressure
before fault with fully charged battery pack.
¢Ó¢Ó Safe backpressure for
8-hours run without fault.
|
|